Current Meter Calibration and Certification
HyQuest Solutions provides calibration and certification services for Mechanical Current Meters. Periodic calibration ensures reliability of measured data. Certificates help you in your QA processes and ensure traceability and auditable documentation. Current… Read more
Current Meter Calibration Tank
Length | 64 meters |
Depth | 1.8 meters |
Starting Velocity | 0.005 m/s |
Time Resolution | 0.001 sec |
Velocity Resolution | ±0.001 m/s |
Width | 2.5 meters |
Maximum Velocity | 8.0 m/s |
Distance Resolution | ± 1.0 mm |
HyQuest Solutions provides calibration and certification services for Mechanical Current Meters. Periodic calibration ensures reliability of measured data. Certificates help you in your QA processes and ensure traceability and auditable documentation.
Current meters are calibrated in accordance with ISO 3455:1976 and ISO 2537:1988.
For quality assurance purposes HyQuest Solutions recommends all Current Meters are calibrated every one year or 300 hours. The maintenance of the current meter consists of replacing the bearings and a full calibration in the high velocity rating tank. All current meters have been calibrated in accordance with AS 3778.6.3-1992. This standard is identical with and has been reproduced from ISO 3455:2007.